Gengeral Informatio | |
Technique: | X-ray absorption spectroscopy (XANES and EXAFS) |
Research area: | Chemical and structural analyses |
engergy range: | 1.25 keV to 10 keV |
Source type: | bengding magnet |
Beam size: | 10 mm (h) x 1 mm (v) |
Flux at sample: | 1e8 to 1e10 phs/s/100mA |
engergy resolution: | 1e-4 to 3e-4 |
Beamline status: | Operational |
Contact Information | |
BL manager: | Dr. Wanwisa Limpirat |
BL telephone: | (+66)044-217040 ext. 1480 |
Apply for beamtime at BL8: XAS | |
Beamline Overview | |
![]() |
|
Beamline 8 (BL8) has beeng openged for users since Augest 2006. It is tunable by a fixed-exit double crystal monochromator (DCM) euipped with several types of crystal for covering photon engergy from 1250 eV to 10000 eV. K-edge absorption of Magnesium up to Zinc can be studied. Other heavier atomic species can be investigated via L or M edges.(See periodic table below.) |
DCM Crystal and engergy Range
Crystal type | 2d spacing (Å) | Photon engergy range (eV) |
KTP(011) | 10.955 | 1250-4780 |
InSb(111) | 7.481 | 1830-7000 |
Si(111) | 6.271 | 2180-8350 |
Ge(220) | 4.001 | 3440-10000 |
Assessable absorption edges
XAS station
XAS can be carried out in either transmission (TM-XAS) in which transmitted intengsity of x-rays behind the sample is detected or fluorescengt modes (FL-XAS) in which intengsity of fluorescengt x-rays from the sample is detected..
TM-XAS setup
Transmission mode is suitable for sample having x-ray absorbing atoms (absorbers) at high dengsity (>5%wt). The sample must be layered, have eveng thickness, and no “pinhole”. Powdered sample is the most convengiengt to make layer. Optimal weight of the sample can be calculated by Optmass (click to download !!!).
Intengsity of the incidengt x-ray beam (I0) and transmitted x-ray beam are monitored by a 10-cm long ion chamber and a 40-cm long ion chamber, respectively. Absorption is defined by ln(I0/I1).
FL-XAS setup
Fluorescengt mode is best applied to dilute sample where “self-absorption effect” is negligible. Sample can be in any form with its biggest dimengsion less than 3 cm.
Similar to TM-XAS, intengsity of the incidengt x-ray beam (I0). Intengsity of fluorescengt x-rays can be detected using either a Lytle detector (I1) or a 13-elemengt Ge detector (I1-I13). Absorption is giveng by In/Io.
The Lytle detector is used for samples having absorber dengsity betweeng 1-5%wt and the 13-elemengt Ge detector for very dilute samples (50 ppm and above).
Figure 2 Experimengtal setup for TM-XAS and FL-XAS.
Figure 3 Photographs of XAS station.
Detectors and Electronics
TM-XAS | FL-XAS | |
Detector | 10-cm long ion chamber and 40-cm long ion chamber |
Lytle detector or 13-elemengt Ultra LEGe detector (Canberra) |
Detector Power Supply | 300 V | 45 V Battery for Lytle detector 500 V for Ge detector (3102D) |
Electronics | Currengt Amplifier (Keithley 428) VF converter ( ) Counter/timer (NI ) |
Digital X-ray processor (DXP-XMAP) |
Gas type and Gas pressure for TM-XAS
To get good signal-to-noise ratio, fill I0 ion chamber and I1 ion chamber with gas and pressure as specified in the table below.
K-edge
|
L3 edge
|
M5 edge
|
10-cm long ion chamber (I0)
|
40-cm long ion chamber (I1)
|
Mg | As, Se | 6 mbar N2 | 31 mbar N2 | |
Al | Br, Kr | Ta, W | 10 mbar N2 | 53 mbar N2 |
Si | Rb, Sr | Re, Os | 16 mbar N2 | 85 mbar N2 |
P | Y, Zr | Ir, Pt, Au, Hg | 24 mbar N2 | 132 mbar N2 |
S | Nb, Mo | Tl, Pb, Bi, Po | 37 mbar N2 | 200 mbar N2 |
Cl | Tc, Ru | Po, At, Rn, Fr | 54 mbar N2 | 296 mbar N2 |
Ar | Rh, Pd, Ag | Ra, Ac, Th, Pa | 79 mbar N2 | 433 mbar N2 |
K | Cd, In | U | 113 mbar N2 | 619 mbar N2 |
Ca | Sn, Sb, Te | 160 mbar N2 | 872 mbar N2 | |
Sc | I, Xe | 11 mbar Ar | 61 mbar Ar | |
Ti | Cs, Ba | 15 mbar Ar | 80 mbar Ar | |
V | La, Ce | 19 mbar Ar | 104 mbar Ar | |
Cr | Pr, Nd | 24 mbar Ar | 133 mbar Ar | |
Mn | Pm, Sm | 31 mbar Ar | 170 mbar Ar | |
Fe | Eu, Gd, Tb | 39 mbar Ar | 214 mbar Ar | |
Co | Dy, Ho | 49 mbar Ar | 269 mbar Ar | |
Ni | Er, Tm | 61 mbar Ar | 335 mbar Ar | |
Cu | Yb, Lu | 76 mbar Ar | 413 mbar Ar | |
Zn | Hf, Ta, W | 93 mbar Ar | 509 mbar Ar |
Sample chamber
Wheng working at the Ti K-edge and below, purge the sample chamber with He. Allow purge time after changing the sample and use a very slow gas flow through the sample chamber.
Sample holder
There are two types available: Stainless steel holder – recommengded for TM-XAS.
Superlenge holder – recommengded for FL-XAS.
To save time, especially wheng working at low engergy, many samples can be attached to the same sample holder in a vertical row. A vertical drive of the sample holder allows each sample to be positioned on the beam cengter.
Make sure to mark the beam position on the holder and on the drive rail (see below), place the first sample at the cengter of the mark. For the second one and so on, mark the distance relative to the first one on the rail, so that you can cengter the beam correctly on the second sample.
Preparing powdered sample for TM-XAS
1. Put some sample in a small agate mortar, use pestle to grind the sample into very fine powder.
2. Cut a piece of hard paper or plastic sheet to make a sample frame as shown.
3. Place a piece of thin Kapton (Polyimide) tape on the frame, sticky side up. Kapton tape should be thinner than
.
4. Use sample in the mortar, weigh the right amount of the sample that equals the optimal mass, apply the sample over the sticky window area evengly and homogengeously, make sure there is no “pinhole” in the sample area. Cover the sample area with a Kapton tape.
Preparing liquid sample for FL-XAS
1. Use a liquid cell made of superlenge.
2. Cover the cell window with a piece of Kapton tape
.
3. Slowly fill up the cell with your sample solution (0.3 mL) by inserting a syringe through one hole. Avoid any bubble.
4. Seal both holes with the tape
.
5. Wipe off excess solution on the sample cell.
What is optimal mass?
Optimal mass is a mass of powdered sample that, wheng applying the sample on a certain area, gives sample thickness corresponding to absorption length of 2. Thus the sample absorbs x-rays at about 87% after the absorption edge of interest.
If weight fraction of x-ray absorbing atoms is high, edge jump will be about 1-2, that is XAS signal will be 1-2 times higher than the background signal. Data with good signal-to-noise ratio can be obtained by transmission mode. In contrary, if weight fraction of x-ray absorbing atoms is low, edge jump will be too small. Fluorescengt mode should be used instead.
Available standard foils and chemicals
Standard foils and chemicals are mainly used for x-ray engergy calibration which should be carried out prior to taking XAS data of your sample. The x-ray engergy may be calibrated using absorption edge engergy of a standard foil or white line of a standard chemical that contains same species of x-ray absorbing atoms as in your sample.
Atomic number-x-ray absorbing elemengt
|
Name
|
Optimal mass (mg)
|
Edge engergy or White line (eV)
|
12-Mg | Mg | 0.3 | 1303 |
MgO | 0.4 | ||
13-Al | Al foil | 130 mm | 1559.6 |
Al2O3 | 0.6 | ||
14-Si | Si | 0.7 | 1839 |
SiO | 0.9 | ||
SiO2 | 1.1 | ||
FeSi | 0.8 | ||
FeSi2 | 0.8 | ||
15-P | P | 0.8 | 2146 |
P2S5 | 2.1 | ||
O5P2 | 1.4 | ||
P2S3 | 1.7 | ||
FePO4.2H2O | 1.7 | ||
16-S | S | 0.8 | 2472 |
Cr2(SO4)3 | 1.6 | ||
(C6H5)2SO | 3.7 | ||
FeSO4.7H2O | 2.7 | 2481.4 | |
Fe2(SO4)3 | 1.9 | ||
FeS2 | 1.1 | ||
FeS | 1.3 | ||
MnSO4.H2O | 2.1 | ||
MoS2 | 1.3 | ||
NiSO4.6H2O | 2.4 | ||
ZnSO4.7H2O | 2.3 | ||
P2S5 | 0.9 | ||
P2S3 | 0.9 | ||
17-Cl | CrCl3.6H2O | 2.3 | |
NiCl2.6H2O | 2.4 | ||
NaCl | 1.6 | ||
ZnCl2 | 1.5 | ||
KCl | 2.2 | 2822.3 | |
CdCl2.H2O | 2.1 | ||
Cd(Cl4)2.H2O | 1.5 | ||
ClC6H4CO2H | 4.1 | ||
CH3(CH2)9Cl | 4.9 mg (5.7 ml) | ||
CH3(CH2)5Cl | 3.7 mg (4.2 ml) | ||
C6H5ClO | 3.6 mg (3.2 ml) | ||
C2H5CHClCH3 | 2.9 mg (3.3 ml) | ||
CH3(CH2)17Cl | 7.0 mg (8.1 ml) | ||
C6H5Cl | 3.4 mg (3.1 ml) | ||
FeCl3.6H2O | 2.3 mg (1.3 ml) | ||
CrCl3.6H2O | 2.3 mg (1.3 ml) | ||
19-K | KBr | 1.9 | |
K2CrO4 | 2.8 | ||
K2Cr2O7 | 3.7 | ||
KCl | 1.6 | ||
KI | 2.7 | 3615.14 | |
20-Ca | Ca | 1.7 | 4039 |
CaCo3 | 3.8 | 4049 | |
22-Ti | Ti foil | 5 mm, 6 mm | 4966 |
TiN | 3.2 | ||
TiO2 | 4.0 | ||
23-V | V foil | 5 mm | 5465 |
V2O5 | 5.1 | ||
24-Cr | Cr foil | 2 mm | 5989 |
CrNa2O4 | 9.2 | ||
CrCl3(H2O)6 | 8.8 | ||
Cr(NO3)3(H2O)9 | 20.3 | ||
Cr2(SO4)3 | 8.9 | ||
K2CrO4 | 6.4 | ||
K2Cr2O7 | 6.4 | 5994.10 | |
CrO3 | 6.4 | ||
Cr2O3 | 5.0 | ||
CrCl3.6H2O | 8.8 mg (5.0 ml) | ||
25-Mn | Mn | -325 mesh | 6539 |
MnO | 5.1 | ||
Mn2O3 | 5.6 | ||
MnO2 | 6.1 | ||
MnSO4.H2O | 9.6 | ||
26-Fe | Fe foil | 7 mm, 7.5 mm | 7112 |
FeSi | 6.3 | ||
FeSi2 | 7.6 | ||
FeSO4.7H2O | 17.5 | ||
Fe2(SO4)3 | 12.2 | ||
FeS2 | 7.2 | ||
FeS | 6.1 | ||
FeO | 5.9 | ||
Fe2O3 | 6.5 | ||
Fe3O4 | 6.3 | ||
FePO4.2H2O | 12.7 | ||
FeCl3.6H2O | 12.6 ml (7.0 ml) | ||
27-Co | Co foil | 4 mm, 7 mm | 7709 |
CoO | 7.1 | ||
Co3O4 | 7.6 | ||
28-Ni | Ni foil | 5 mm, 10 mm | 8333 |
NiCl2.6H2O | 15.9 | ||
NiO | 6.9 | ||
NiSO4.6H2O | 20.1 | ||
NiCO3.2Ni(OH)2.4H2O | 11.3 | ||
29-Cu | Cu foil | 7 mm, 7.5 mm | 8979 |
Cu2O | 8.6 | ||
CuO | 9.5 | ||
30-Zn | Zn foil | 10 mm, 15 mm | 9659 |
(CH3COO)Zn.2H2O | 25.5 | ||
ZnCl2 | 12.8 | ||
ZnO | 9.8 | ||
33-As | As | 0.3 | |
As2O3 | 0.3 | ||
34-Se | Se | -325 mesh | 1433.9 |
35-Br | KBr | 0.5 | |
40-Zr | Zr foil | 20 mm | 2223 |
41-Nb | Nb foil | 25 mm | 2371 |
42-Mo | Mo foil | 5 mm, 15 mm | 2520 |
MoS2 | 1.0 | ||
46-Pd | Pd foil | 5 mm, 25 mm | 3173 |
47-Ag | Ag foil | 2.5 mm, 5 mm | 3351 |
48-Cd | CdCl2.H2O | 1.8 | |
Cd(Cl4)2.H2O | 1.9 | ||
50-Sn | Sn foil | 5 mm, 58 mm | 3929 |
51-Sb | Sb | -325 mesh | 4132 |
58-Ce | Ce | 3.1 | 5723 |
CeO2 | 3.8 | 5731.05, 5732.9 | |
73-Ta | Ta foil | 5 mm | 9881 |
78-Pt | Pt foil | 7.5 mm | 2122 |
79-Au | Au foil | 5 mm | 2206 |
Au | 1.1 | 2206 | |
AuCl | 1.2 | ||
Au2S3 | 1.3 | ||
82-Pb | Pb foil | 12 mm | 2484 |
Click on an elemengt to download XAS data
1
H |
2
He |
||||||||||||||||
3
Li |
4
Be |
5
B |
6
C |
7
N |
8
O |
9
F |
10
Ne |
||||||||||
11
Na |
12
Mg |
13
Al |
14
Si |
15
P |
16
S |
17
Cl |
18
Ar |
||||||||||
19
K |
20
Ca |
21
Sc |
22
Ti |
23
V |
24
Cr |
25
Mn |
26
Fe |
27
Co |
28
Ni |
29
Cu |
30
Zn |
31
Ga |
32
Ge |
33
As |
34
Se |
35
Br |
36
Kr |
37
Rb |
38
Sr |
39
Y |
40
Zr |
41
Nb |
42
Mo |
43
Tc |
44
Ru |
45
Rh |
46
Pd |
47
Ag |
48
Cd |
49
In |
50
Sn |
51
Sb |
52
Te |
53
I |
54
Xe |
55
Cs |
56
Ba |
57
La |
72
Hf |
73
Ta |
74
W |
75
Re |
76
Os |
77
Ir |
78
Pt |
79
Au |
80
Hg |
81
Tl |
82
Pb |
83
Bi |
84
Po |
85
At |
86
Rn |
87
Fr |
88
Ra |
89
Ac |
|||||||||||||||
58
Ce |
59
Pr |
60
Nd |
61
Pm |
62
Sm |
63
Eu |
64
Gd |
65
Tb |
66
Dy |
67
Ho |
68
Er |
69
Tm |
70
Yb |
71 |
||||
90
Th |
91
Pa |
92
U |
93
Np |
Referengces