Experimental Station

The SUT-NANOTEC-SLRI XAS Experimental Station is designed to accommodate a wide variety of samples, focusing on determining local atomic structures and oxidation states.

Standard Setup

  • Sample Stage: A versatile mounting stage adaptable for standard holders or specialized in-situ chambers.
  • Transmission Mode: Ionization chambers before and after the sample measure X-ray flux absorption. This mode is suited to concentrated samples.
  • Fluorescence Mode: A four-element Silicon Drift Detector is placed perpendicular to the beam path for dilute samples or thin films.
Figure 3 Schematic of the SUT-NANOTEC-SLRI beamline station
Figure 3 Schematic of the SUT-NANOTEC-SLRI beamline station
Experiment mode Detector and beam size
Transmission Detector: 10 mm and 40 mm ionization chambers
Beam size at sample: 13 mm (W) × 1 mm (H)
Fluorescence 4-element Silicon Drift Detector, Vortex-ME4 model 267-VTX-ME4
Sensitive thickness: 0.5 mm
Active area: 50 mm2 each, 200 mm2 total
Beam size at sample: 20 mm (W) × 1 mm (H)

Absorption Edges Available at BL5.2

Figure 4 Absorption edges that can be measured at BL5.2
Figure 4 Absorption edges that can be measured at BL5.2