Experimental Station
The primary advantage of the TRXAS end-station is its ability to record complete X-ray absorption spectra simultaneously and rapidly. This makes it an ideal tool for in-situ measurements to monitor structural and electronic changes in samples under various environmental conditions, such as high temperature, pressure, or specific gas atmospheres.
In-situ Environmental System
Currently, the institute provides a specialized Heating Cell designed to support a wide range of in-situ research applications, enabling simultaneous temperature control and atmospheric regulation.

Technical Specifications

Maximum Operating
Maximum Operating
Temperature:
973 K (700 °C)
Ramp Rate:
5-20 °C/min
Controller:
Eurotherm 2216e
with K-type
thermocouple
Stability:
± 0.1 K
Gas Atmosphere
& Flow Control
The cell is designed to accommodate various gas
environments for catalytic and material study.
Available Gases:
N2, O2, He, Ar, CO2 and H2
Flow Control:
AALBORG Mass Flow
Controllers (MFCs)
Flow Rate
Ranges:
0–100 sccm and
0–25 sccm
Safety & Operational Requirements
Gas Handling: Use of flammable gases such as H2 requires prior safety clearance andspecific experimental protocols.
Cooling: The water-cooling system must be active when operating at temperatures
above 473 K (200 °C) to protect the Kapton windows and seals.
In-situ Battery System


Coin Cell Analyzer 5V10mA Battery Testing System

Technical Specifications
AC Input
AC 220V±10% 50Hz
Power
25 W
Resolution
AD: 16bit; DA: 12bit
Input
≥1MΩ
Voltage
Measuring Range
Discharge Min Voltage
Accuracy
Stability
Charge:25mV~5V
Discharge: 25mV~5V
0V
± 0.1% of FS
± 0.1% of FS
Current
Range
Accuracy
Stability
Charge: 0.05mA~10mA
Discharge: 0.05mA~10mA
± 0.1% of FS
± 0.1% of FS
Power
Output Power Per Channel
Stability
0.05W
± 0.2% of FS
Time
Rise Time
Step Time
1ms (0~Full Range)
≤ (365*24) hour/step
Time Format 00:00:00 (h:min:s)
Charge
Mode of Operation
End Conditions
CCC, CVC, CC & CVC
VOltage, Current, Test Time, Capacity
Discharge
Mode of Operation
End Conditions
CCD, CPD
Voltage, Current, Test Time, Capacity
Cycle
Cycles
Steps
Nested Function
65535
63
Max three levels of loops
Detector System
To measure XAS spectrum, the position sensitive detector (PSD) is used.

S3904–1024F
NMOS Linear Image Sensor
with Scintillator
(Hamamatsu Photonics K.K.)
NMOS linear image sensor consisting of 1024 pixels of an N-type diffusion layer formed on a P-type silicon substrate.


Detection Method
The use of an indirect detection method, employing an X-ray scintillator (150 μm thick CsI (Cesium Iodide) coated on a fiber optic plate) to convert X-rays into visible light, extended the sensor lifetime to over a year.

Absorption Edges Accessible at BL2.2
Click on each element with a icon to view the available standard at BL2.2